![Agilent Technologies 34410A Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/agilent-technologies/34410a/34410a_user-manual_2868275075.webp)
34410A/11A User’s Guide
75
Features and Functions
2
The following tables list the automatic trigger delays for ac measurements,
triggered from the remote interface. From the front panel, assume an
automatic trigger delay of 50 ms for all ac measurements.
AC Voltage Measurements (all ranges):
AC Current Measurements (all ranges):
Frequency and Period
Temperature measurements actually use the 2–wire or 4–wire resistance
functions. The above tables for resistance apply to temperature
measurements, but all temperature measurements use autoranging. Since
the range in use cannot be predicted, nor can the delay. Use the
TRIGger:DELay?
command to show the actual trigger delay.
Capacitance measurements have a trigger delay of zero, because
discharging the capacitor is part of the measurement time.
AC Filter
Trigger Delay
Slow (3 Hz)
2.5 sec
Medium (20 Hz)
625 ms
Fast (200 Hz)
25 ms
AC Filter
Trigger Delay
Slow (3 Hz)
1.66 sec
Medium (20 Hz)
250 ms
Fast (200 Hz)
25 ms
Trigger Delay
1 sec
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...