20
Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
Making a Basic State Measurement
The logic analyzer automatically triggers on
Pattern1
, the first trigger
event. You can set up more complex triggers by editing the sequence levels
and combining trigger events.
a. Select the sequence level number
1
button and choose
Edit...
b. In the dialog box, select the
Pattern1
button just after Trigger on and
choose
Combo...
c. In the Combination box, select the
Off
option button next to
Pattern2
and choose
On
.
d. To change the trigger to
Pattern1 and Pattern2
, select the
Or
option
button to the right of the events and choose
And
.
e. Select the
OK
button.
f.
Select the
Close
button.
The analyzer is now set to trigger when it detects both the pattern
defined by
Pattern1
and the pattern defined by
Pattern2
on the target
system's buses. The trigger sequence windows shows
Trigger on "(Pattern1.Pattern2)" 1 time
See Also
“The Trigger Tab” on page 73
1. Select the
Run
button.
2. Examine the data.
a. Select the
Window
menu.
b. Select
Analyzer<A>
in the menu and choose
Listing<1>
.
Depending on what other instruments are active, there may be more
than one
Analyzer<A>
. Choose the one that refers to your analyzer.
c. To have the listing display appear automatically when you run the logic
analyzer, select Options -> Popup on Run -> On in the menu bar of the
listing display.
d. To insert additional labels, select the label name.
See Also
For Connection Information