What is a Calibration Procedure
Calibration procedures verify that products or systems operate within
the specifications. Parameters covered by specifications have a
corresponding calibration procedure. Calibration procedures include
both performance tests and system verification procedure. Calibration
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance, which is
referred to as an "operational accuracy calibration".
What is a Function Test
Function tests are quick tests designed to verify basic operation of a
product. Function tests include operator's checks and operation
verification procedures. An operator's check is normally a fast test
used to verify basic operation of a product. An operation verification
procedure verifies some, but not all, specifications, and often at a lower
confidence level than a calibration procedure.
Agilent Technologies 16712A Logic Analyzer
Specifications
The specifications are the performance standards against which the
product is tested. These specifications apply only to the Agilent
Technologies 16712A 128K Sample logic analyzer:
Maximum State Speed: 100 MHz
Threshold Accuracy: +/- (100mV + 3% of threshold setting)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com