RN200 Mono Software User Manual
14
Parameter name
Description
Setting range
Power meter Address
Communication address of the power
meter
Attenuation Value
External Attenuator's attenuation value
Frequency Start
Starting frequency point for calibration
24000-44000MHz
Frequency Stop
Stopping frequency point for calibration
24000-44000MHz
Frequency Step
Frequency step for calibration
Minimum 0.01MHz
Trace Name(S-Parameter)
Calibration path
S21
IF Bandwidth
Test bandwidth
Table 4: Emission calibration parameters
3.4.1.2.2 Emission Calibration
First set up a calibration connection, as shown in Figure 17. A power meter and an attenuator
are required. Connect the attenuator to the transmission port of the positioner with a RF
cable. When the calibration connection is correctly completed, click "Start Calibration."