RN200 Mono Software User Manual
33
Figure 32: Template Import Window
As shown in Figure 32, click the "Import" button in the parameter template window to open
the test report list. The user can choose any completed 3D EIRP Scan test report from this
list, and view parameter configurations and test results of the report. Click the "OK" button,
and the system will automatically apply the selected test results and its parameters to the
parameter template of the current test.
When selecting the “3D EIRP Scan” option, the “3D EIRP Scan” test will be performed first.
The system will automatically calculate the optimal positioner test position and optimal
Beam ID according to the test results, then proceed with the following tests.