RN200 Mono Software User Manual
21
Each parameter’s definition and setting range is listed in Table 5. After setting all the
parameters correctly, click "Next" to start the Passive/Active_Tx calibration.
Parameter name
Description
Setting range
Frequency Start
Starting frequency point for calibration
24000-44000MHz
Frequency Stop
Stopping frequency point for calibration
24000-44000MHz
Frequency Step
Frequency step for calibration
Minimum 0.01MHz
Trace Name(S-Parameter)
(
link to read data
)
Calibration path
S21
IF Bandwidth
Test bandwidth
Passive Power
transmit power
Power range in the
Emission file
Standard Gain File
Gain value data of a standard gain antenna
Table 5: Calibration Parameters
3.4.1.2.4 Passive/Active_Tx Calibration
Before the Passive/Active_Tx calibration, connect a standard gain antenna to the transmit
port of the positioner with a RF cable, place the antenna perpendicular to the Phi surface of
the positioner, and close the chamber door.