GPIB OPERATION
3-2-37
SWEEP OPERATION
IMODE
Select Instrument Mode
Data type :
Character Program Data (either NORMAL for signal generator
operation or SWEEPER for swept operation )
Allowed suffices :
None
Default suffix :
None
Example:
IMODE SWEEPER
SWEEP
[not used alone]
:MKRON
Enable Sweep Markers
:MKROFF
Disable sweep Markers
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples:
SWEEP:CFRQ:START 75MHZ;STOP 150MHZ;STEP 100;TIME 10MS
SWEEP:RFLV:START -56DBM;STOP -12DBM;STEP 440;TIME 25MS
SWEEP:CFRQ:MKRNUM 1;VALUE 83MHZ;MKRON
:CFRQ
[not used alone]
:<cmd>
Select Carrier Frequency sweep parameter entry where <cmd>
is replaced by one of the commands (START, STOP, STEP,
TIME, MKRNUM, MKRON, MKROFF or VALUE)
:RFLV
[not used alone]
:<cmd>
Select RF Level sweep parameter entry where <cmd> is
replaced by one of the commands (START, STOP, STEP,
TIME, MKRNUM, MKRON, MKROFF or VALUE)
:LFGF
[not used alone]
:<cmd>
Select LF Generator Frequency sweep parameter entry where
<cmd> is replaced by one of the commands (START, STOP,
STEP, TIME, MKRNUM, MKRON, MKROFF or VALUE)
:LFGL
[not used alone]
:<cmd>
Select LF Generator Level sweep parameter entry where
<cmd> is replaced by one of the commands (START, STOP,
STEP, TIME, MKRNUM, MKRON, MKROFF or VALUE)
:INTF
[not used alone]
:<cmd>
HOP
:<cmd>
Select Internal Modulation Oscillator Frequency sweep
parameter entry where <cmd> is replaced by one of the
commands (START,
STOP, STEP, TIME, MKRNUM,
MKRON, MKROFF or VALUE)
[not used alone]
Select Frequency Hopping sweep parameter entry where
<cmd> is replaced by TIME.
:START
Select start value of the parameter to be swept.
:STOP
Select stop value of the parameter to be swept.
Data type :
As used for the parameter
Allowed suffices :
As used for the parameter
Default suffix :
As used for the parameter