GPIB OPERATION
3-2-42
MISCELLANEOUS COMMANDS
IMODE
Select Instrument Mode
Data type :
Character Program Data (either NORMAL for signal generator
operation or SWEEPER for swept operation )
Allowed suffices :
None
Default suffix :
None
RPPR
Reset reverse power protection
trip
Data type :
None
Allowed suffices :
None
Default suffix :
None
FSTD
Select internal or external frequency standard
Data type :
Character program data (any one of INT0, INT1, INT5, INT10,
EXT1, EXT5 or EXT10)
Allowed suffices :
None
Default suffix :
None
Examples:
:FSTD INT10
:FSTD EXT5
FSTD?
Prepares message containing information on frequency
standard selection in the format:
:FSTD <char>
Example:
:FSTD EXT10
BLANK
Blank or unblank various parts of the display. The number sent
after the command determines the action to be taken as
follows:
0 blank or unblank the Carrier Frequency display
1 blank or unblank the RF Level display
2 blank or unblank the Modulation Frequency display
3 blank or unblank the Modulation display
4 blank all displays
Data type :
Decimal Numeric Program Data (any one of 0, 1, 2, 3 or 4)
Allowed suffices :
None
Default suffix :
None
Examples:
BLANK 0
BLANK 4
BACKL
[not used alone]
:ON
Backlighting On
:OFF
Backlighting Off
Data type :
None
Allowed suffices :
None
Default suffix :
None
Examples:
BACKL:ON
BACKL:OFF