ACCEPTANCE TESTING
5-1-29
Oscilloscope
Volts/div 0.2
V
Time base
5 ms/div
Spectrum analyzer
Preset and calibrate
Reference frequency
108.1 MHz
Reference level
10 dBm
Span/div 50
Hz
Resolution bandwidth
10 Hz
(3)
The typical traces on the oscilloscope and the spectrum analyzer are shown in
Figs. 5-1-16 and 5-1-17. These traces are for the 90 Hz and the 150 Hz tones'
having equal amplitudes (i.e. 0% DDM).
Fig. 5-1-16 Oscilloscope trace for a 0% DDM waveform.
(4)
On the UUT reduce the SDM depth using the rotary control. This will cause a
reduction in the amplitude of the oscilloscope trace and the sidebands on the
spectrum analyzer trace. Reset the SDM to 40%.
Fig. 5-1-17 Spectrum analyzer trace for a 0% DDM waveform.
C0526
Ch. 1
= 200.0 mVolts/div
Timebase = 5.00 ms/div