ACCEPTANCE TESTING
5-1-10
Modulation oscillator distortion
Fig. 5-1-5 Modulation oscillator distortion test set-up
(1)
Connect the test equipment as shown in Fig. 5-1-5.
(2)
Set the UUT to
[Source Freq: F4]
100 Hz.
(3)
Check that the distortion measured on the audio analyzer at the frequencies
indicated in Table 5-1-6 is less than 0.1%.
TABLE 5-1-6 DISTORTION FREQUENCY SETTINGS
Modulation oscillator frequencies
100 Hz
10 kHz
1 kHz
20 kHz
C0219
LF
OUTPUT
UUT
Audio Analyzer
INPUT
➤