ACCEPTANCE TESTING
5-1-19
FM scale shape
(1)
With the test equipment connected as in Fig. 5-1-9, set the UUT to
[Carrier Freq.]
31.64 MHz
[RF Level]
0 dBm
[FM Devn.]
224 kHz.
(2)
Referring to Table 5-1-14, measure the FM on the modulation meter at the
deviations indicated.
TABLE 5-1-14 FM SCALE SHAPE DEVIATIONS
Deviation (kHz)
224 256 289
228 261 293
233 265 298
238 270 302
242 275 307
247 279 312
252 284 316