![Advanced Instruments OsmoPRO Multi-Sample Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/advanced-instruments/osmopro-multi-sample/osmopro-multi-sample_user-manual_2856047076.webp)
OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
76
A - Product specifications
Sample type
Aqueous based solution
Units of measure
mOsm/kg H
2
O
Sample volume
20 µL
Sample capacity
Up to 20 samples
Test time
90 seconds (typical) to test single sample
40 minutes (typical) to test 20 samples
Power requirement
100 to 240 VAC (50/60 Hz) / 0.5 A
Power consumption 60W max.
Fuses (2): 250V time delay (Type T) 1-Amp
Memory backup
Integral lithium cell; 10-year life (typical); (not user-replaceable)
Classifications
Protection class: I
Over-voltage category: II
Pollution degree: 2
Moisture protection: IPX0 (ordinary)
Communications
USB 2.0 Type A ports (3)
USB 2.0 Type B ports (1)
Ethernet port (1)
Operating conditions
Temperature: 18 °C to 35 °C (64 °F to 95 °F)
Humidity: 30 to 80% relative humidity (non-condensing)
Storage temperature
-40 °C to 45 °C (-40 °F to 113 °F)
Dimensions (D x W x H)
37 cm x 25 cm x 44 cm (14.5” x 10” x 17.5”)
Net weight
13.2 kg (29 lbs.)
Shipping weight
19.1 kg (42 lbs.)
Warranty
One-year limited warranty on workmanship and parts
Certifications