OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
10
1. Installation & setup
WARNING:
It is important to read and follow the steps in this section. Read Chapter 2
System Overview before attempting to run tests on the instrument.
1.1 Find a location for the instrument
When choosing a location for the Advanced Instruments OsmoPRO Multi-Sample Micro-
Osmometer, be sure to meet the following criteria:
Adequate space: The dimensions of the instrument are 12” Wide x 13” Deep x 17” High (31 cm
x 33 cm x 43 cm). Be sure to keep the workplace free of debris and allow 6 inches (15 cm) of
clear space on both sides and behind the instrument for fan driven air circulation.
Electric outlet availability: The instrument needs to operate within five feet of a properly
grounded, three-prong electrical outlet capable of continuously supplying 2 amperes at
100-240VAC 50/60Hz. If the instrument is not grounded properly, its operation may be
impaired and it may be a safety hazard. Therefore, be sure to test the outlet and record the
results before operating the instrument.
IN THE CASE OF EMERGENCY:
Power to the instrument can be removed by pulling the
detachable power cord from the electrical outlet.
1.2 Unpack the instrument
To unpack the instrument, take the following steps:
•
Carefully unpack the instrument, accessories and supplies, and inspect them for
shipping damage. Use the enclosed packing list (Table 1.1) to verify all items were
received.
•
Save the shipping boxes and packaging material in case future transport is necessary.
•
If any item on the packing list appears to be missing from the shipment, search carefully
through and under all packing materials. If the item is not found, notify your receiving
department immediately. Advanced Instruments is only responsible for items reported
missing within 10 days of a shipment’s arrival.
•
If any damaged items are received, save the cartons and packing material in which
those items came for inspection by the insurer. The carrier, dealer, and Advanced
Instruments must be notified within 24 hours for the warranty and insurance to apply.
Have the transportation company inspect items, fill out a “Report of Concealed
Damage,” and file a claim. Notify Advanced Instruments immediately for repair or
replacement.
Complete the online warranty card to register your product.