Owner’s Guide
54
Separation Distances
•
The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled.
•
Portable and mobile RF communications equipment should be used no closer
to any part of the device, including cables, than the recommended separation
distance in the table below.
•
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey, should be less than the compliance level in each frequency range.
Rated maximum output
power of transmitter, P
Separation distance
150 kHz to 800 MHz
d = 1.17√P
800 MHz to 2.5 GHz
d = 2.33√P
0.01 W
0.1 m
0.2 m
0.1 W
0.4 m
0.7 m
1 W
1.2 m
2.3 m
10 W
3.7 m
7.4 m
100 W
11.7 m
23.4 m