Owner’s Guide
53
Electrode type:
Suitable for metal microelectrodes (300 kΩ typical)
Dimensions:
8 mm OD, 100 mm long
Material:
Stainless steel
Cable length:
2.8 m
This equipment is not intended to be modified or serviced by the user. No user serviceable
parts inside. Refer servicing to authorised ADInstruments service centre.
ADInstruments
reserves the right to alter these specifications at any time.
Electromagnetic Compatibility
The FE285 Neuro Amp EX (the device) has been tested to comply with the requirements
of IEC 60601-1-2, (IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-
4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-8, IEC 61000-4-11 and CISPR 11).
Emissions
•
The device uses RF energy for its internal function only. RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
•
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Immunity
•
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the device requires continued operation during
power mains interruptions, it is recommended that the device be powered from
an uninterruptible power supply or a battery.
•
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
•
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Содержание BIO
Страница 1: ...Owner s Guide BIO AND NEURO AMPLIFIERS ...
Страница 18: ...Owner s Guide 16 ...