Owner’s Guide
22
Applied parts
Single/Dual Bio Amp:
•
MLA2540 (Dual) Bio Amp Cable
•
MLA2505 Lead wires
•
MLA1212 Micro-Hook Electrodes (Single)
Octal/Quad Bio Amp:
•
MLA0310 Lead wires, unshielded
Other Supplied Accessories
Single/Dual Bio Amp:
•
MLAC01 BNC-BNC cables
•
MLAC02 DB9M- DB9M cable
•
MLA2350 (Single) Bio Amp Cable
Octal/Quad Bio Amp:
•
MLAC27 Quad BNC-BNC cables
•
MLAC02 DB9M- DB9M cable
The Bio Amp Cable
Connections should be made to the Bio Amp and Dual Bio Amp inputs using the supplied
Bio Amp cables and leads. The Dual Bio Amp cable plugs into the six-pin input socket on
the Dual Bio front panel: a notch in the plug ensures that polarity is correct. The lead
wires on the Bio Amp cable plug into the input sockets on the Bio Amp front panel. Only
the supplied Bio Amp cable and leads should be used. Other cables may not meet safety
requirements.
NEG
RA
(AHA)
LA
RL
REF
POS
To Bio Amp
To Dual Bio Amp
To electrodes
To electrodes
Applied parts of
the Bio Amps
Figure 3–3
Bio Amp cable
yokes, with leads
attached: 3 leads
for the Bio Amp,
and 5 leads for
the Dual Bio Amp
Содержание BIO
Страница 1: ...Owner s Guide BIO AND NEURO AMPLIFIERS ...
Страница 18: ...Owner s Guide 16 ...