8230E Optical Power Meter Operation Manual
4.6.2 Remote Control Command List
4-19
System
Instrument infor-
mation
*IDN?
Query command which inquires for the instru-
ment information.
New model name response
ADC Corp.,8230E ,XXXXXXXXX,YYYYY
ADC Corp.: Manufacturer (9 characters)
8230E : Instrument name (5 characters)
XXXXXXXXX: Serial number (9 characters)
YYYYY: ROM revision number (5 characters)
The new model name response is applied to
software revision D00 or later.
Old model name response
ADC,8230E ,XXXXXXXXX,YYYYY
ADC: Manufacturer (3 characters)
8230E : Instrument name (5 characters)
XXXXXXXXX: Serial number (9 characters)
YYYYY: ROM revision number (5 characters)
Information for
options *5
*OPT?
Query command which inquires for the instru-
ment option information
Reply: No option
0: (1 character)
Option 70
+70: (3 characters)
User
parameter
SA0
Saves the setting parameters in area [0] of the
non-volatile memory.
SA1
Saves the setting parameters in area [1] of the
non-volatile memory.
SA2
Saves the setting parameters in area [2] of the
non-volatile memory.
SA3
Saves the setting parameters in area [3] of the
non-volatile memory.
CL
Saves the values when shipping in areas [0] to
[3] of the non-volatile memory.
RC0
Loads the values, which are saved in area [0] of
the non-volatile memory, as the setting parame-
ters.
RC1
Loads the values, which are saved in area [1] of
the non-volatile memory, as the setting parame-
ters.
RC2
Loads the values, which are saved in area [2] of
the non-volatile memory, as the setting parame-
ters.
RC3
Loads the values, which are saved in area [3] of
the non-volatile memory, as the setting parame-
ters.
RL
Loads the values when shipping as the setting
parameters. *3
*3:
MYID and the key lock status are not changed.
*5:
This command is applied to software revision C00 or later.
Table 4-5 Remote Control Commands (5 of 7)
Item
Command
Description
Initial value
Availability
Power
ON
Factory
default
Calibration
mode
Error2