8230E Optical Power Meter Operation Manual
7.6.2 Current Measurement Calibration
7-9
200 nA :
XR5
2
A :
XR6
20
A :
XR7
200
A :
XR8
2 mA :
XR9
20 mA :
XR10
50 mA :
XR11
4. Open the input (0A).
5. Perform the zero calibration.
Zero calibration
XPC0
6. If an error occurs, initialize the range and re-execute calibration.
7. Perform the zero calibration in all ranges in the order of steps 3 to 5 shown above.
Perform the full calibration in all ranges.
8. Set the range.
9. Connect the VIG, refer to Table 7-2 Current Measurement Calibration, and output
the calibration point value.
10. Input the VIG generation value (the value that is determined by the national stan-
dard) and perform the calibration.
Entering the value XPCfull (Input the determined value in “full.”)
11. If an error occurs, initialize the range and re-execute the zero and full calibration.
12. Perform the full calibration in all ranges in the order of steps 8 to 11 shown above.
13. When the zero and full calibrations are complete in all ranges, obtain the zero and
full calibration data and ensure that the data is correct.
14. If an abnormality occurs, initialize the abnormal range and re-execute the zero
and full calibrations.