8230E Optical Power Meter Operation Manual
7.5 Calibration Operation
7-5
7.5
Calibration Operation
Calibration of this instrument cannot be performed by using the key pad. Perform the calibration by remote
control through the USB. For more information on USB remote commands, refer to calibration items in Sec-
tion 4.6.2, “Remote Control Command List”.
There are two types of calibration: The IV offset calibration and the current measurement calibration. The IV
offset calibration must be complete before performing the current measurement calibration.
Figure 7-3 Calibration Flow
Calibration start
Prepare the register
and cable
IV offset calibration
Prepare the VIG and cable
Current measurement
calibration
End of calibration