6.0 CFC USER CALIBRATION
•
The
CFC
scales calibrate using metric weights and the
CFCa
scales can calibrate using either metric or pound weights,
depending on the weighing unit in use before calibration.
•
When the parameter menu shows "
F1 CAL
" press the
[Tare]
key.
•
The display will then show "
unLoAd
" requesting for all weights
to be removed from the platform and press the
[Tare]
key.
•
The display will then show the last calibration weight
requested.
•
If this weight is acceptable press the
[Tare]
key to accept it or
press
[CE]
to enter a new calibration weight using the keypad
and then press the
[Tare]
key. To calibrate at 100.00 Kg
simply type 10000. Do not type the decimal place.
•
The display will show "
LOAd
". Place the calibration weight on
the scale and press the
[Tare]
key. The scale will calibrate
and then start the initial self-check procedure counting from 9
to 0. Remove the weight during the counting.
CFCa
scales will also have the lb or kg indicator
ON
to show the denomination of the
weight requested. If the scale was in pounds before starting calibration, the weights
requested will be in pound values or if the scale was weighing in kilograms then
metric weights will be requested.
•
If an error message “
FAiL L
” or “
FAiL H
” is shown in the
“
WEIGHT
” display during calibration, repeat the calibration. If
the problem persists then contact Adam Equipment or your
dealer for further assistance.
•
After calibration, you should check whether the calibration
and the linearity are correct. If necessary repeat calibration,
ensure that the scale is stable before accepting any weight.