F6 Prt
Sets the RS-232. Press
[Tare]
to show the following sub-menus.
P bAUd-
Select from the desired baud rates
Press
[Tare]
to see the earlier setting. Press
[M+]
to scroll through
the options
(
600, 1200, 2400, 4800
and
9600)
. Press
[Tare]
to
select the desired value and return to the sub-menu.
P nodE -
Sets the printing mode to-
AU on
(Print anytime a weight is placed on the scale).
AU oFF
(Print whenever the
[Print]
key is pressed).
P Cont
(Print continuously
the unit weight, no. of pieces and the
total weight).
SEI rE
(Continuously print the weight only)
Press
[Tare]
to select and return to the sub-menu.
PArity –
Select from the following settings-
n 8 1 -
8 data bits, no parity
E 7 1 -
7 data bits, even parity
o 7 1
-
7 data bits, odd parity
Press
[Tare]
to select and return to the sub-menu.
Press
[Zero]
to return to the menu.
F7 SPd
This is used to set the speed at which the scale will run the ADC. The settings
available are 7.5, 15, 30 and 60- the slowest setting being 7.5 and the fastest
is 60. Press
[Tare]
to select the desired value and return to the menu. Press
[Zero]
to return to weighing.
NOTE:
If the User Parameters cannot be accessed using the default
code “0000”, use the factory code. See section 7.0 for details on Factory
parameters.