3.0 ERROR
CODES
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than the
permissible value (typically 4%
of maximum capacity) when
power is turned on or when the
[Zero]
key is pressed,
Weight on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing the scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard error.
Improper operation of the
scale.
Err 6
A/D count is not correct when
turning the scale on.
Platform is not installed.
Load cell may be damaged.
Electronics may be
damaged.
FAIL H
or
FAIL L
Calibration error
Improper calibration.
If the problem persists
contact your dealer or Adam
Equipment for assistance.
@Adam Equipment Company 2007
5
Содержание CFC SERIES
Страница 2: ...Adam Equipment Company 2007...
Страница 4: ...Adam Equipment Company 2007 2...
Страница 19: ...Adam Equipment Company 2007 17...