The software will look for the electrode with switchid=0 and switchaddress=2. The id
of the remote electrode (for instance102) must be unique within the spread and not 0.
12.6
Spread Files for Pole-pole
Two remote electrodes must be specified in the spread file.
<Cable>
<!-- Add remote electrodes -->
<Name> LS Panel </Name>
<Electrode>
<Id> 102 </Id>
<Name> C2 Current </Name>
<SwitchId> 0 </SwitchId>
<SwitchAddress> 2</SwitchAddress>
</Electrode>
<Electrode>
<Id> 104 </Id>
<Name> P2 Potential </Name>
<SwitchId> 0 </SwitchId>
<SwitchAddress> 4</SwitchAddress>
</Electrode>
</Cable>
The id 102 and 104 for C2 and P2 can be selected arbitrary. In protocol itself, 0 can be
used as electrode number. In case of extension of the system with additional electrode
selectors (ES10-64) there will be real electrodes in with id 101, 102, 103, 104 etc.,
and then other arbitrary id numbers such as for example 501, 502, 503, 504 can be
used for remote electrodes. Here it is only important to have a unique id for each
electrode. The software uses the switch id and switch address to find the remote
electrode.
12.7
Protocol Files for Pole-pole
Example of protocol file
<Measure>
<Tx> 1 0 </Tx>
<Rx>
49 0
<Focus>
<X> 25 </X>
<Z> 41.66 </Z>
</Focus>
</Rx>
The line "<Tx> 1 0 </Tx>" will tell the software to translate the reference to
electrode id 0 to the "C2 Current" electrode, and <Rx> 49 0 will tell the software to
translate the reference to electrode id 0 to the "P2 Potential" electrode.