90
Linearity
0.005 %
Range
Auto range of measurement
±2.5V, >100M
Ω
±15V, 30 M
Ω
(fully differential)
±600V, 20 M
Ω
Flat Frequency Response
better than 1 % up to 300 Hz
10.4
Transmitter
Output power
250 W
Current transmission
True Current Transmitter
Output Current Accuracy
better than 0.4 %
Maximum Output Current
2500 mA
Maximum Output Voltage
+ / - 600 V (1200 V peak to peak)
Instant Polarity Changer
YES
Accuracy
0.4 %
Precision
0.1 %
Self Diagnostics
Temperature, Power dissipation monitoring
Safety
Emergency Interrupter easily accessible
Tx Monitor
Full waveform monitored
Accuracy
0.2 %
Precision
0.1 %
10.5
Relay Switch
Switching matrix
Internal 10 X 64
Roll-along
YES full coverage, both 2 & 3D. All 84 take-
outs in an ABEM standard four-cable array are
active for roll-along
Array types Default
Multiple gradient, dipole-dipole, pole-dipole,
pole-pole, Wenner, Schlumberger, etc.
Electrode Test
YES, Focus One
10.6
Software & Communication
Terrameter LS is controlled using the incorporated Firmware. It supports Multi-
Electrode Survey Systems for 2D & 3D for Resistivity, IP & SP Imaging &
Monitoring.
It has a Graphical User Interface that is easy to follow in all its aspects. Clear and
instructive graphics assists the user in the operation of the instrument.
For enhanced Data Quality Control in the field it is possible to display the measured
Multi-Electrode Resistivity Imaging data in near real time as a pseudo section. Thanks
to access points as USB and RJ 45 (for LAN) transfer of data to other computers is
extremely simple.
For full inversion of data external software is required. Most common today is
RES2DINVor RES3DINV. This program supports data formats provided with the
help of the Terrameter LS software.