Figure 41. IP settings view
The timing setup for the off-time measurements in IP-mode is defined in the
"IP
Window Settings" View
(Figure 41). The current-off time will be equivalent to "IP
Minimum Off Time" except if the total sum of IP delay times and the sum of
integration time for IP windows are larger. It is recommended to set each IP time
window to a multiple of the local power line net frequency period time (for instance
20, 40, 60, or 100 ms in case of 50 Hz frequency).