ABEM Terrameter LS
98
<Focus>
<X> 52 </X>
<Z> 3.192 </Z>
</Focus>
</Rx>
</Measure>
Note that the number of channels in an instrument does not restrict the number of
potential measurements for a current electrode pair, and that the same measurement
sequence file can be used for an instrument with 4, 8 or 12 channels
6
.
12.4
Spread Files for Pole-dipole
It is important that the spread file contains the remote electrodes that are in use.
Adding the example part of the spread file shown below can do this.
Example of a part of the spread file for pole-dipole:
<Cable>
<!-- Add remote electrodes -->
<Name> LS Panel </Name>
<Electrode>
<Id> 102 </Id>
<Name> C2 Current </Name>
<SwitchId> 0 </SwitchId>
<SwitchAddress> 2</SwitchAddress>
</Electrode>
</Cable>
12.5
Protocol Files for Pole-dipole
Example of the protocol file
<Measure>
<Tx> 23 0 </Tx>
<Rx>
22 21
<Focus>
<X> 22.25 </X>
<Z> 0.52 </Z>
</Focus>
</Rx>
</Measure>
The line "<Tx> 23 0 </Tx>" will tell the software to translate the reference to
electrode id 0 to the "C2 Current" electrode.
6
Note: The overall efficiency in the use of the measuring channels may differ and it might be possible
to optimise the array combinations differently depending on the number of the measuring channels.