12.3 Appendix III – Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The CentriMag Circulatory Support System with 2
nd
Generation CentriMag Console and Mag Monitor is intended for
use in the electromagnetic environment specified below. The customer or the user of the CentriMag Circulatory
Support System should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
Group 1
The System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR11
Class A
The System is suitable for use in all establishments other than
domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies