Circulatory Support System Operation Manual (US)
Page 26 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
CAUTION
The Monitor is intended for use only with the 2
nd
Generation CentriMag Console.
WARNING
The Monitor can only be operated when it is connected to the 2
nd
Generation
CentriMag Console. The Monitor may be operational on AC power or, in
combination with the Console Version 2, on battery. Operating the Monitor while
the Console runs on battery shortens the battery runtime. Refer to the individual
RVAD or LVAD Console display for all operational data and audio/visual alarm
messages.
WARNING
Only USB-compatible Memory Sticks may be used to connect to the USB Port of
the Monitor. No other USB device may be used with the USB port (e.g. printer).
5.1.5 2
nd
Generation CentriMag Primary Console
The Console uses single-phase AC power when used with the CentriMag Pump. The
Pump is capable of a flow rate of up to 10.0 LPM or maximum pressure head of 600
mmHg. Flows up to 1.5 LPM or a maximum pressure head of 540 mmHg may be
generated with the PediMag Pump and circuit. In addition, each Console contains a
rechargeable internal battery that is capable of maintaining Console functionality in the
event of a loss of AC power.
5.1.5.1 Front Panel
The Console (
) is a microprocessor-based device. The
microprocessors generate the primary Motor control signal. The Monitor sensors generate
front display outputs and provide alarm functions. The microprocessors acquire the sensor
data for use in generating operator displays and alarms. A graphical screen is used to
display monitored data, System options, and menus. Operator settable alarms and
parameters are accessible via the System menus.