Table 16: Console Alarms & Alerts
ID
Alarm/
Alert
Text Message
System Status & Operator Response
B3
Alert
BATTERY CHARGER FAIL
Press the alarm acknowledge button. If the alert message
repeats, switch to backup System as described in Section 10.1.
If this alarm is associated with BATTERY MODULE FAIL then
carry out the procedure associated with that alarm.
B4
Alert
BATTERY MAINTENANCE
REQUIRED
Do not use the Console. Perform battery maintenance
according to the instructions provided in Section 9.4.
B5
Alert
LOW BATTERY
Plug the Console into AC power outlet to charge battery.
If no AC outlet is available, switch to backup Console, Motor
and Flow Probe according to the procedure described in
Section 10.1.
Resume support.
B6
Alert
ON BATTERY
Verify that the user wants the Console to be on battery. If so,
carefully monitor the status of the battery charge indicator,
while using the System on battery.
Re-connect to AC-outlet, as soon as possible.