CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 121 of 126
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
EM-TEC PEDIATRIC FLOW PROBE FOR PEDIMAG
TM
PUMP
(AVAILABLE SEPARATELY)
PARAMETER
SPECIFICATIONS
Product Name
em-tec Pediatric Flow Probe
Design
Ultrasonic transit time technology
Physical Specification
Width (L1)
33mm (1.29“)
Length (L2)
45mm (1.77“)
Thickness (L3)
25mm (0.98“)
Accuracy
0.0 to 1.0 LPM
± 0.1 LPM + Offset Drift
1.0 to 8.0 LPM
±7 % of the value + Offset Drift
Maximum Slope Error
max. 0.03 LPM within 2 hours
Ultrasound Frequency
15 kHz to 18 MHz, different patterns
possible, resolution 8 bit
Resolution
1 mLPM
Retrograde Flow Detection
Measured to -2.0 LPM
Tubing Specification
Tubing ID
6.4 mm (1/4”)
Wall Thickness
2.4 mm (3/32”)
Tubing OD
11.1 mm (7/16”)
Material
Polyvinyl Chloride
Probe Calibration
Calibrated for Fluid
Temperature (ºC)
Blood at 37 ºC
Tubing Type
Tygon S-50-HL