CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 85 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
2) Reboot:
This menu item allows users to exit the Management Application and return to normal
operation, Note: This will restart the Monitor only and not the Console.
Once “Reboot” is selected, press the
UP
key and then the
DOWN
key (confirm) to reboot
the Monitor.
The Information Panel on the right displays the following information:
1) Actual Time/Date Setting
2) Application Selection Setting: Always “CentriMag”
3) Debug Mode Setting: Always “Off”
4) Data Logger View Mode Setting: Always “Normal”
5) Application Software Version
7.26 Pump Set-up
Refer to the CentriMag and PediMag Pump Instructions for Use for proper setup and
operation of the Pumps.
WARNING
Always fully unscrew the Pump retaining screw built into the Motor before
inserting and locking the Pump in the Motor receptacle. This requires five
complete counter-clockwise rotations of the screw. Failure to do so may inhibit
the ability to fully seat and lock the Pump in the Motor receptacle resulting in
loss of function and a MOTOR ALARM or PUMP NOT INSERTED alarm. Should
this condition occur, unscrew the retaining screw, remove the Pump, reinsert
the Pump, tighten the retaining screw, turn the Console power OFF and ON,
ensure no alerts/alarms are displayed, and set the Console to initiate pumping.