In the case of multiple alarms/alerts, each time a new alarm/alert condition occurs, a new
audible alarm and a new message will be displayed on the Console and Monitor. The
alarm/alert messages will be listed in order of priority from most severe to less severe.
Alarms always take higher priority than alerts. If more than three alarm/alert conditions
occur simultaneously, the Console will offer the DOWN option. Depressing the
DOWN
keypad will scroll down one alarm/alert message. Refer to
for
a listing of alarm
and alert messages in their order of priority.
The Console features 6 alarms and 20 alerts as shown in
Table 11: Console Alarms & Alerts In Order Of Priority
ID
Alarm/
Alert
Description
Ability to Silence Audio
(Yes/No)
(Silence Interval
)
S1
Alarm
POWER ON TEST FAIL
No
S2
Alarm
SYSTEM FAULT (Run-Time System
Failure)
No
M1
Alarm
MOTOR STOPPED
Yes (60 Sec.)
M2
Alarm
MOTOR DISCONNECTED
Yes (60 Sec.)
M3
Alarm
PUMP NOT INSERTED
Yes (60 Sec.)
M4
Alarm
MOTOR ALARM
No
M5
Alert
SET PUMP SPEED NOT REACHED
Yes (60 Sec.)
B1
Alert
BATTERY MODULE FAIL
Yes (60 Sec.)
B2
Alert
BATTERY BELOW MINIMUM
Not while running on
batteries – can be silenced
when reconnected to AC
F1
Alert
FLOW PROBE DISCONNECTED
Yes (60 Sec.)
S3
Alert
SYSTEM ALERT
Yes (60 Sec.)
F2
Alert
FLOW SIGNAL INTERRUPTED
Yes (60 Sec.)
F3
Alert
FLOW BELOW MINIMUM
Yes (60 Sec.)
F4
Alert
FLOW ABOVE MAXIMUM
Yes (60 Sec.)
P1
Alert
PRESSURE 1 DISCONNECTED
Yes (Permanent – visual
message remains)
P2
Alert
PRESSURE 2 DISCONNECTED
Yes (Permanent – visual
message remains)
8
When applicable, audio tone will reactivate if the condition persists during time lapse.