CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 34 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
6.2 Specifications
Table 5:
Console Specifications
PARAMETER
SPECIFICATIONS
AC Power
100 – 240 VAC
at 50/60 Hz, 170 VA
Battery Power
14.8 VDC Li-Ion, internal rechargeable battery
Discharge time
: approx. 120 minutes @3,500 RPM,
5.5 LPM
Recharge time
: 4 hrs to 90% charge, 5 hrs to 100%
charge
Dimensions
Height: 10.0 cm / 3.9 in
Width: 26.6 cm / 10.5 in
Depth: 33.0 cm / 13.0 in
Weight
5.8 kg / 12.8 lbs
Pump Speed Range
0 – 5,500 revolutions per minute (RPM)
Pump Flow Range
0.0 – 10.0 liters per minute (LPM)
Flow Range Display
-2.0 – 10.0 liters per minute (LPM)
Electrical Safety
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
6.3 Environmental Conditions
6.3.1 Shipping Conditions
The following are the acceptable environmental conditions during shipping:
•
Temperature: -29
o
C to 60
o
C (one week maximum)
•
Relative humidity: 0% to 85%
•
Atmospheric pressure: 210 hPa – 1100 hPa (157 mmHg – 825 mmHg)
6.3.2 Operational and Storage Conditions
The following are the acceptable environmental conditions during operation and storage:
•
Temperature: 10
o
C to 30
o
C
•
Relative humidity: 30% to 75%
•
Atmospheric pressure: 702 hPa – 1100 hPa (527 mmHg – 825 mmHg)
1
Note: If the probe detects forward flow of more than 10 LPM then it will display as “^^.^^ LPM”.