Distance for cardiac pacemakers = Minimum clearance from the primary loop area to be
observed by persons wearing cardiac pacemakers.
2
)
This value is also valid for smaller primary loops with two windings.
Fig. 1.5: Minimum clearances to primary loops for persons wearing cardiac pacemakers (principle representation)
Marking for the
minimum clearance
Top view
3D representation
2CDC352016F0003.EPS
1000 - 1200
8.0
2
)
800
1200 - 1400
10.0
1000
1400 - 1600
12.0
1300
1600 - 1800
13.0
1400
1800 - 2000
15.0
1600
2000 - 2200
16.0
1700
2200 - 2400
18.0
1900
2400 - 2600
20.0
2100
2600 - 2800
22.0
2300
2800 - 3000
24.0
2500
Short
edge of the
primary loop [mm]
Current [A]
Distance for cardiac
pacemakers [mm]
1
)
Long
edge of the primary loop (one winding):
1200 – 6000 mm
Minimum clearance to
the primary loops for
persons wearing cardiac
pacemakers