Component
Order number
Fuse (T 500 mA, 250 V) for power
supply, suitable for all devices
3KQR000443U0100
10 . . .Repair
Replacing the LCD indicator
G12030
1
3
4
2
3
4
1
5
A
B
Integral mount design
Remote mount design
Housing cover
Fixing screws for LCD indicator
LCD indicator
Connector
Fixing screws for housing cover (4x)
A
B
1
2
3
4
5
Figure 46 Replace LCD indicator
Replace the LCD indicator as follows:
1
Switch off the power supply.
2
Unscrew / remove the cover.
3
Loosen the screws for the LCD indicator (only on integral
mount design).
4
Remove the LCD indicator.
5
Pull the connector out of the electronic.
6
Plug in the connector of the new LCD indicator. Ensure
that the cable harness is not damaged.
7
Insert the LCD indicator and screw on /replace the cover.
8
Switch on the power supply.
Note
If the O-ring gasket is seated incorrectly or damaged, this
may have an adverse effect on the housing protection
class .
Follow the instructions in chapter "Opening and closing the
terminal box" on page 14 to open and close the housing
safely.
The LCD indicator can be replaced in the event of a
malfunction.
Component
Ordering number
LCD indicator (HMI).
Applicable for integral and remote mount
design.
3KQZ407125U0100