7 Operation
Safety instructions
Menu navigation
Risk of burns due to hot measuring media
The device surface temperature may exceed 158 °F (70 °C),
depending on the measuring medium temperature.
Before starting work on the device, make sure that it has
cooled sufficiently.
CAUTION
Aggressive or abrasive measuring medium may result in
damage to the wetted parts of the flowmeter sensor. As a
result, pressurized measuring medium may escape
prematurely.
Due to wear on the flange seal or process connection gaskets
(e.g., threaded pipe connections, Tri-Clamp, etc.), a
pressurized measuring medium may escape.
When using internal flat gaskets, these can become
embrittled through CIP/SIP processes.
If there is a chance that safe operation is no longer possible,
take the device out of operation and secure it against
unintended startup.
The LCD indicator has capacitive operating buttons. These
enable you to control the device through the closed housing
cover.
Note
The transmitter automatically calibrates the capacitive
buttons on a regular basis. If the cover is opened during
operation, the sensitivity of the buttons is firstly increased to
enable operating errors to occur. The button sensitivity will
return to normal during the next automatic calibration.
You can use the
or
operating buttons to browse
through the menu or select a number or character within a
parameter value.
Different functions can be assigned to the
and
operating buttons. The function that is currently assigned to
them is shown on the LCD display.
Control button functions
Menu
Exit
Select
2
3
4
5
5
Operating buttons for menu navigation
Menu name display
Menu number display
Marker for indicating relative position within the menu
Display showing the current functions of the
and
operating buttons
1
2
3
4
5
Figure 43 LCD display
Meaning
Exit
Exit menu
Back
Go back one submenu
Cancel
Cancel a parameter entry
Next
Select the next position for entering numerical and
alphanumeric values
Meaning
Select
Select submenu / parameter
Edit
Edit parameter
OK
Save parameter entered