1-6
2000-OSM, F1
Gas Control (Digital)
Control Method:
Closed loop PID; temperature stabilized
Number of Zones:
1 to 5
Filtration:
2 microns at inlet, provided
Inlet Pressure:
Minimum:
Se 10 psig (69 kPa, 0.7 bar)
Maximum:
100 psig (689 kPa, 6.89 bar)
NOTE
It is the customer's responsibility to set relief devices used to protect the analyzer
and prevent hazardous releases at a maximum of 200 psig (1380 kPa, 13.80 bar).
Range:
0 to 100 psig (0 to 689 kPa, 0 to 6.89 bar), bubble tight, non-
venting
Gauges:
Electronic readout: 0.01 psig (0.07 kPa, 0.7 mbar) resolution
Setpoint resolution: 0.01 psig (0.07 kPa, 0.7 mbar)
Temp Coefficient:
None
Regulation:
See below
Accuracy:
0 to 50 psig:
1.7%
50 to 100 psig:
2.7%
Repeatability:
±0.1 psig (±0.7 kPa, ±7 mbar)
Allowable Gasses:
H
2
, He, N
2
, Air, Ar
Quality:
GC
Grade
Flow Adjustment:
Oven mounted needle valves or flow controllers with external
adjustment
Tube Fittings:
316 SS Gyrolok (standard)
316
SS
Swagelok
(optional)
1/16,
1/8,
1/4
inch
connections