2000-OSM, F1
4-20
4. On the second Print Tables screen, enter the Method Table number and then press the F2 (Print
Selected Table) soft key. The Method Table and Component Table will print (see Figure 4-19).
METHOD TABLE #01 8:25:29 Fri 30 Sep 1994
Cycle Time = 00140
TIME
FUNCTION VALUE
0001
Noise
Calc
On
0003
Noise
Calc
Off
0004
Auto
Zero
N/A
0005
Stream
Step
N/A
0010
Temp
Check
One
0015
Slope
Detect
On
Now
0016
Pres
Check
One
0017
Valve
On
One
0019
Valve
Off
One
0028
Component
One
0037
Valve
On
Two
0039
Valve
Off
Two
0048
Component
Two
0049
Pres
Control
One
0057
Valve
On
One
0059
Valve
Off
One
0062
Temp
Control
Two
0068
Component
Three
0077
Valve
On
Two
0079
Valve
Off
Two
0088
Component
Four
0097
Valve
On
One
0099
Valve
Off
One
0108
Component
Five
0138
Slope
Detect
Off
Now
0139
Component
Six
0139
Component
Seven
0139
Component
Eight
0139
Component
Nine
METHOD TABLE #01 8:25:57 Fri 30 Sep 1994
NAME
RRT
W
KEY
LL
HL
RF
CC
METHANE 28.00 5.0
Y 0.00 100. 30.673
95.00
%
ETHANE 48.00 5.0
N
0.00 100. 0.50051
0.9689
%
ETHYLENE
68.00 5.0 N 0.00 100. 0.16836
0.5210
%
H2
88.00
5.0
N
0.00 100. 1.7311
3.351
%
CO
108.0
5.0
N
0.00 100. 0.051707
0.1599
%
Z
139.0
5.0
N
0.00
100. 1.0000
0.0000
Clc
REAL HEAT
139.0
5.0
N
0.00 50000.
1.0000
100.0 Clc
SORT OF B
139.0
5.0
N
0.00 100. 1.0000
100.0
Clc
IDEAL HEAT
139.0
5.0
N
0.00 9000.
1.0000
100.0
Clc
Figure 4-19. PRINTED METHOD TABLE
5. When all the data have printed, press the F1 soft key as necessary to return to the Background
screen.
PEAK DETECTION
The analyzer measures peaks in two different ways: forced gating and slope detection. Both
methods can exist in the same analysis, but they cannot both measure peaks at the same time. If the