8-12
2000-OSM, F1
30. Install two
cone point set screws
into the
cylinder
and tighten into the groove on the
body
.
31. Perform “Installing an LSV on the Analyzer.”
Installing the LSV on the Analyzer
1. Open the isothermal oven door.
2. Install
the
vaporizer assembly
around the
vaporizer chamber
.
3. Install the four screws that retain the two parts of the
vaporizer assembly
.
4. Install
the
insulation
around the
vaporizer assembly
.
5. Install
the
insulation cover
over the
vaporizer assembly
and
insulation
.
6. Install the LSV on the analyzer.
7. Orient
the
flange
so that the marks made at LSV removal are lined up. This should have the
sample chamber
vertical with the air connections at the top of the valve and the connection of
the
vaporizer chamber
at the top.
8. Install the four screws holding the
flange
to the side of the analyzer and tighten the screws.
9. Reconnect all tubing on the Liquid Sample Valve.
10. Close the isothermal oven door.
Replacing the LSV Vaporizer Assembly
1. On the Manual Temperature Control Mode screen, note which zone has the LSV connection.
2. Stop the analysis at the end of a cycle.
3. Turn off power, carrier, sample, and air to the analyzer and allow it to cool.
4. Open the Controller front and side doors, and the isothermal oven door.
5. In the Digital Temperature Controller, locate the Liquid Sample Valve’s Zone Board.
6. Disconnect the LSV heater and sensor wires, which make up the vaporizer cable.
7. Withdraw the free end of the vaporizer cable from beneath the Card Cage into the Controller
housing.
8. Inside the isothermal oven, loosen any cable clamps securing the vaporizer cable assembly and
free the cable.
9. Label all tubing connections to the LSV.
10. Disconnect all tubing connections to the LSV.
11. Mark the analyzer and
flange
to ensure the
flange
is reinstalled in the correct orientation.
12. Remove the four screws holding the
flange
to the side of the analyzer.
13. Separate the LSV from the analyzer.