2000-OSM, F1
7-19
15. Perform FID cell maintenance for cleaning jet and flame out thermocouple (see “FID
REPAIR/Cleaning the FID Jet” in Section 8).
16. Replace the plastic shield, cover, and thumb nuts on the Detector Amplifier.
17. Close and tighten the Controller front door.
TP 30: CHECKING TEMPERATURE TABLE VALUES
Verify that the software and hardware temperature parameter values are correct and have not been
inadvertently changed. To perform this operation, it is necessary to access the temperature
configuration table and the manual temperature control mode.
1. Press the F1 soft key until the Background screen is displayed (see Figure 7-3).
**********
B A C K G R O U N D
**********
ABB Analytical GCC
Remote 0.1
803V029-1R (Std GCC)
(C) 2001
**ALARM**
**ALARM** 18-JUN-2002 TUE 08:16:01
**ALARM**
Analysis Time
:00114 Peaks Found:00
Cycle Time
:00000 State:Idle
Current Stream
:01
Key Component:
Purging Stream
:01 none
Method
:01 entered
Exit
Manual
Cur Report,
To
Control
Basic I/O & Alarms
Commands
Options
F1
F2
F3
F4
Figure 7-3. BACKGROUND SCREEN
2. On the Background screen press the F2 (Manual Control) soft key.