![ABB C310 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/abb/c310/c310_programming-manual_3052157018.webp)
15
4.5
Set Up Process Variable
Information.
•
Universal inputs – mV, mA, V, THC, RTD and resistance.
•
Internal cold junction compensation.
•
Linearization of temperature sensors to allow use of non-linearizing transmitters or any
electrical input.
•
Programmable fault levels and actions.
•
Digital filter reduces the effect of noise on inputs.
4
CONFIGURATION…
Input
Type
Linearizer
Type
Linearizer
Units
Linearizer
Range
Display Range
5/2
3/2
√
RTD
THC N
THC B
THC E
THC J
THC T
THC S
THC R
THC K
None
Programme
Filter
°
F
°
C
RTD
THC
Current
Voltage
Millivolts
Display Range
High
Value set to 0
Value set low
Value set high
Fault
Detection
Level 10 %
200
0
Display Range
Low
Fault
Detection
Level 10 %
200
220
–20
0
Example – Type K thermocouple Range 0 to 200
°
F with 10% fault detection levels.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com