CALIBRATION
…7.3
Factory Settings
Factory Settings Access Code
Enter the required code number (between 0000 and 19999) to gain access to the factory
settings. If an incorrect value is entered, access to subsequent frames is prevented and the
display reverts to the top of the page.
Calibrate Sensor A
Note.
The values in the display lines for sensor calibration are shown as examples only –
the actual values obtained will differ.
Sensor B calibration (dual input analyzers only) is identical to Sensor A
calibration.
See page 60.
Operating Page
– see Section 2.3.
Calibrate Input for Sensor A ?
If calibration is required select
Yes
otherwise select
No
.
Note.
To abort calibration, press the
key again at any time before calibration is
complete – see next page.
Millivolt Zero
Set the millivolt source to –1000mV.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note.
The upper 6-segment display shows the measured input voltage. Once the signal
is within range the lower 6-segment display shows the same value and
Calib
is displayed
to indicate that calibration is in progress.
Millivolt Span
Set the millivolt source to +1000mV
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (100R)
Set the temperature simulator to 100
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A:mV Zero (-1V)
-1000
Calib
A:mV Span (+1V)
mV
Calib
A:T.Zero (100R)
Ohms
Calib
Cal. Sensor B
Monitoring pH
-1000
mV
A: T.Zero (1K0)
1000
1000
100.0
100.0
Cal. Output 1
Calibrate I/P ?
-----
Yes
No
Yes
No
Monitoring Redox
Dual Redox
pH/Redox