12
8.30
pH
A1: Setpoint
Sen.A
Temp.A
VIEW SETPOINTS
-----
35.0
Deg.C
A2: Setpoint
Sen.B
6.80
A3: Setpoint
Temp.B
55.0
Deg.C
A4: Setpoint
-----
Off
A5: Setpoint
VIEW OUTPUTS
SENSOR CAL.
Security Code
CONFIG. DISPLAY
VIEW SETPOINTS
pH
View Set Points
This page shows alarm set points. The value of each of the set points is shown, together
with the name of the parameter it is assigned to.
Alarm assignments, set point values and relay/LED actions are programmable – see
Section 5.4.
Sensor A (pH), Alarm 1 Set Point
Sensor A (Temperature), Alarm 2 Set Point
Sensor B (pH), Alarm 3 Set Point
– Dual input analyzers only
Sensor B (Temperature), Alarm 4 Set Point
– Dual input analyzers only
Note.
Alarm 4 available only if option board fitted
and
analog features enabled – see
Section 7.3.
Alarm 5 Set Point
Note.
Alarm 5 available only if option board fitted
and
analog features enabled – see
Section 7.3.
See Section 3.2.
See Note on next page.
3
OPERATOR VIEWS
3.1
View Set Points
Note.
The parameter names and units of measurement displayed in the
View Set Points
page depend on the
Probe Type
settings for Sensors A and B – see Section 5.3. Those shown below are given as examples only.