16
…3
OPERATOR VIEWS
The logbook stores data entries for alarm events, sensor errors, power failures and pH
calibration information.
View Logbook
Use the
and
keys to access the
Alarms
logbook.
Note.
If no entries are stored in the
Alarms
logbook, the display shows
No More Entries
.
Alarms
The
Alarms
logbook contains up to 10 entries (entry 1 is the most recent), each comprising
an alarm number, alarm state (On or Off) and the date/time of the occurrence.
Option board fitted
and
analog features enabled (Section 7.3) – see Section 3.6.
See Note on Page 13.
Advance to entries 2 to 10.
Note.
If no more entries are stored, the display shows
No More Entries
.
View Logbook
Use the
and
keys to access the
Errors
logbook.
Note.
If no entries are stored in the
Errors
logbook, the display shows
No More Entries
.
Errors
The
Errors
logbook contains up to 5 entries (entry 1 is the most recent), each comprising
the sensor letter, error number and the date/time of the occurrence.
Option board fitted
and
analog features enabled (Section 7.3) – see Section 3.6.
See Note on Page 13.
Advance to entries 2 to 5.
Note.
If no more entries are stored, the display shows
No More Entries
.
-----
-----
11:09:02 12:34
VIEW CLOCK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Sen.A
VIEW LOGBOOK
Errors
1
Sen.A
Pt100
2
Alarms
Cals
Power
Errors
-----
11:09:02 12:34
VIEW LOGBOOK
-----
VIEW CLOCK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
A1
-----
VIEW LOGBOOK
1
2
Cals
Power
Errors
Alarms
A1
On
3.5
View Logbook
Note.
The View Logbook function is available only if the option board is fitted
and
analog features enabled (Section 7.3)
and
Logbook
is set to
On
(Section 5.9).