Sensor Calibration
…4.1.4
Automatic, Single- and Two-Point Calibration (pH Only)
…4
SETUP
Calibrate Buffer 2 (Two-Point Calibration only)
Immerse Sensor A in the second buffer solution.
Press the
key to initiate calibration.
Note.
To abort calibration, press the
key again at any time before calibration is
complete – see previous page.
The center display line shows the measured sensor output in millivolts.
As calibration procedes, a progress indicator appears in the lower display line. When the
measured sensor output stabilizes, the lower display line shows
##### 100% #####
.
The display then changes for 2 seconds to show the temperature-corrected buffer value in
the upper display line, then advances automatically to the next frame.
Calibration Message
See Table 4.1 for details of calibration messages.
Slope Value
% slope value.
A value between the programmed minimum % slope value (see
Set Min Slope
in the
CONFIG. SENSORS
page – Section 5.3) and 105% is displayed. If the value is outside these
limits, check the electrode system.
pH check value.
Displayed as an additional indication of electrode system condition; 7pH is the optimum
value for glass electrodes and 0pH for Antimony electrodes.
Sensor B (dual input analyzers only) calibration is identical to Sensor A
calibration.
Probe Type
for Sensor B (dual input analyzers only) set to
Redox
or
ORP
and
Enable Cals.
set to
No
(Section 5.3) – return to top of page.
Alter Sec. Code
not set to zero (Section 5.8) – see Section 5.1.
Alter Sec. Code
set to zero (Section 5.8) – see Section 5.2.
SECURITY CODE
CONFIG. DISPLAY
Sensor Cal. A
A: Immerse Buf.2
9.18
pH
20.0
Deg.C
#### 100% ####
128
mV
A: Calibration
-----
Passed
A: Slope & Check
100.0
%
7.00
pH
A: Cal. Buffer 2
4.01
pH
A: Buffer Method
set to
Auto 1-Pt
A: Buffer Method
set to
Auto 2-Pt
Sensor Cal. B
SENSOR CAL.