33
5
PROGRAMMING…
Config. Sensor A
-----
A: pH Glass
-----
A: Ref. Checking
-----
CONFIG. DIAGS
-----
Off
On
Off
On
A: Ref. Alarm
50
kOhm
CONFIG. ALARMS
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B
Config. Sensor B
Config. Sensor B
Config. Sensor A
On
Off
5.4
Configure Diagnostics
Note.
The Configure Diagnostics function is applicable only if
Diff. Input
for Sensor A and/or Sensor B is set to
Yes
– see
Section 5.3.
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
pH Glass Check
Set to
On
to enable the impedance of the glass electrode circuit to be monitored in order
to provide low glass impedance and out-of-sample/broken cable warnings.
Note.
This frame is displayed only if
A: Probe Type
is set to
pH
A: Electrode
is set to
Glass
– see Section 5.3.
Reference Electrode Check
Set to
On
to enable the condition of the reference electrode to be monitored in order to
provide:
– an indication of when sensor cleaning or replacement is required
– out-of-sample/broken cable warnings
Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
See Section 5.5.
Reference Alarm
Set the impedence value above which the reference alarm is activated.
Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
See Section 5.5.