61
7
CALIBRATION
Calibrate Output 3
Notes.
• Output 3 (and Output 4) calibration is applicable only if the Option board is fitted
and
analog features enabled – see below.
• Output 3 calibration is identical to Output 2 calibration.
Calibrate Output 4
Note.
Output 4 calibration is identical to Output 3 calibration.
Configure Option Board
Notes.
• This frame is displayed only if an option board is fitted.
• The software detects if an option board is fitted but cannot detect the additional features
available.
• If an option board is fitted, the correct selection must be made below to enable use of the
available features. If an incorrect selection is made, the software menus and frames
associated with that option are displayed in the Operating and Configuration pages but
the features do not work.
Use the
and
keys to enable the features for the type of option(s) fitted:
Analog
– Analog features enabled (comprising two additional analog
outputs, two additional alarm relays, clock and logbook facility).
Pb Dp
– PROFIBUS-DP digital communications features enabled.
Pb Dp
– Both analog and PROFIBUS-DP features enabled.
Alter Factory Code
Set the factory settings access code to a value between 0000 and 19999.
Return to main menu.
Operating Page
– see Section 2.3.
…7.3
Factory Settings
Cal. Output 3
-----
Cal. Output 4
-----
Option Board
Analog
-----
Pb Dp
Alter Fact. Code
0000
Monitoring pH
FACTORY SETTINGS
Option Board
Fitted
and
Analog
Features Enabled
Option Board
Not
Fitted
Option Board Fitted,
Additional Features
Disabled
Redox (ORP)
Dual Redox
pH/Redox (ORP)