4-4
Circuits
Bridge
Configurations
Examples
Bridge Box Wiring
Methods
Remarks
One-gauge
configuration
-Suited for where simple tension,
compression, or bending force
is applied
-Suited for where changes in
ambient temperature are small
-Calculated using the calibration
value as it is
One-gauge,
three- wire
configuration
-Suited for where simple tension,
compression, or bending force
is applied
-Strain-gauge-lead wires are
temperature-compensated.
-Calculated using the calibration
value as it is
one-active
gauge,
one-dummy
gauge
configuration
-Suited for where simple tension,
compression. or bending force
is applied
-Temperature compensation
using a dummy gauge
-Calculated using the calibration
value as it is
Two-active
Gauge
configuration
-Suited for where simple tension,
compression, or bending force
is applied
-Temperature compensation
-Calculated using calibrated
value x 1/ (1+
). or
signal value x 1/ (1+
)
Two-active
Gauge
configuration
-Detects bending strain
-Eliminates tension and
compression strain
-Temperature compensation
-Calculated using calibration
value x 1/2 or signal value x 1/2
Table 4-2 Wheatstone Bridge Connections (1)
Содержание AS2503
Страница 1: ...AS2503 AS2603 DC BRIDGE SYSTEM DYNAMIC STRAIN AMPLIFIERS INSTRUCTION MANUAL 1WMPD4003356 ...
Страница 2: ......
Страница 4: ......
Страница 30: ......
Страница 46: ......
Страница 68: ......
Страница 83: ...9 5 9 3 External Dimensions 9 3 1 Amplifier Unit AS2503 ...
Страница 84: ...9 6 9 3 2 Amplifier Unit AS2603 ...
Страница 85: ...9 7 9 3 3 Panel Cut Sizes ...
Страница 87: ...9 9 9 3 5 Rack mount Case Rack Rail Elongate Hole ...
Страница 88: ...9 10 9 3 6 Bridge Box 5370 and 5373 ...