3-11
3.6 Cases
3.6.1 Names of Case Elements
Fig.3-7
・
Use standard AC power cord (47326) that is attached to the case as a standard for 100 VAC
power supply.
・
Use optional DC power cord (47229) for DC power supply.
3.6.2 How to Use Case
A) Power supply cable connection
When using a battery (12 VDC, or 10 to 30 VDC) for power supply with amplifier units being
installed in a case, take into account the voltage drop caused by the DC power cord length and
wire diameter. When using several channels or a long power supply cord, a voltage drop occurs.
This voltage drop may cause the power supply voltage to be lower than the allowable power
supply voltage of 10 V at DC power input connector.
For example, the DC power supply cord (47229) has a core area of 1.25 mm
2
. If eight
amplifiers are mounted, a current of 3.2 A (0.4 A x 8) flows, and a voltage drop of 0.5 V will occur if
the cord is extended to 10 m. If a 10-m cable with 0.75 mm
2
is used, a voltage drop of 1.65 V will
occur. For use under such conditions, the power supply should be provided with the voltage drop
taken into account, or the wire diameter or cord length of the power cord should be modified.
B) Synchronization between units and cases
When several units are mounted in the case, wires for signals are connected inside of the case.
BAL, calibration value application, or key lock can be made for all units in the case using the BAL
switch, calibration application switch, or key lock switch for all channels positioned on the front
panel.
Moreover, the synchronization is also available by connecting cases using the synchronization
cable (Figure 3-8).
AC Power Input
Ground terminal
Interface Connector
DC Power Fuse
AC Power Fuse
DC Power Input
Содержание AS2503
Страница 1: ...AS2503 AS2603 DC BRIDGE SYSTEM DYNAMIC STRAIN AMPLIFIERS INSTRUCTION MANUAL 1WMPD4003356 ...
Страница 2: ......
Страница 4: ......
Страница 30: ......
Страница 46: ......
Страница 68: ......
Страница 83: ...9 5 9 3 External Dimensions 9 3 1 Amplifier Unit AS2503 ...
Страница 84: ...9 6 9 3 2 Amplifier Unit AS2603 ...
Страница 85: ...9 7 9 3 3 Panel Cut Sizes ...
Страница 87: ...9 9 9 3 5 Rack mount Case Rack Rail Elongate Hole ...
Страница 88: ...9 10 9 3 6 Bridge Box 5370 and 5373 ...