3-14
3.6.3 Heat Release for Case
A) Using standalone rack-mounting case
As a rack-mounting case does not have legs, avoid placing it on a desk or floor. Otherwise, heat
cannot be released, resulting in damages to amplifiers. It should be properly mounted.
Figure 3-11 Rack-mount Case Installation
B) Using two or more rack-mounting cases
In this case, install fans following the criteria below because the temperature in the unit rises
depending on the number of stacks in the rack, load, and ambient temperature.
Floor or Desk
Rack-mounting case
20mm
and
over
CAUTION
Table 3-7 Number of Fan for Heat Release and Rack-mount
Case
Number of case
Number of fan unit
under severe condition
1-3
1
3-6
2
6-9
3
Note
What are harsh environmental conditions?
- Power supply: 110 VAC(+10%)
- Output voltage and current: +10V, 10mA
- Ambient temperature: +50
C